Chlorine Sensor GCM
The total chlorine sensor GCM serves
to determine the total chlorine concentration of potable
water, pool water or water of comparable quality. The
sensor is mounted to the sampling station DCM 01.
The TOPAX NT serves as measuring and control unit. Only
by using the sampling station DCM 01 the desired inflow
parameters are achieved which are necessary for a safe
operation of the sensor. Air bubbles in the sample water
may cause false measuring results and have to be avoided.
The sampling station DCM 01 must be operated in a pressureless
condition. For monitoring the flow a flowmeter is provided
which opens a contact when no flow is present. This
contact sets the control of the connected TOPAX NT to
0 % thus avoiding undefined metering. The total chlorine
sensor has to be serviced and checked for impurities,
plant contamination and air bubbles regularly. Impurities
can be removed by means of a water jet.
|