TOPAX® MV 450

Instruction Manual

General
TOPAX® MV 450-type measuring amplifiers are analog devices for the measurement of values important in water treatment, i.e.: pH value, free chlorine, REDOX value and conductivity. A 3 1/2- digit, 13 mm LC display of the amplifier shows the actual value. The standard TOPAX® MV 450 are equipped with two relays for alarm signaling (max. and min. alarm). The alarms can be set continuously over the whole measuring range. If the set value is not reached (min.) or exceeded (max.), the corresponding relay is activated. The voltage-free make-and-break contacts are connected to the terminal strips of the device. Red LEDs on the front plate indicate an alarm.

After connection of the power supply the measuring amplifier is ready for operation. For calibration of
the measurement inputs, adjusting controllers are available. After opening the transparent cover, the
zero point and the range can be set at the relevant potentiometer by using a screw driver.

Control elements and displays
Digital display
The digital display shows the currently measured value or the limit values. During normal operation the currently measured value is displayed. To set and display the limit values, the tipping switch must be pushed to the left (min. limit) or to the right (max. limit).

In one of these positions, the required value can be set by means of a screw driver. If the amplifier is used for pH value or REDOX measurement and no sensor is connected or there is a failure, a malfunction of the display is possible,

i.e., except for "1" in the first place no digits are visible. In this case the connector plug, the cable or
the combination electrode must be checked. By short-circuiting the BNC plug, the display is set to
7.00 pH for the pH value and 0 mV for the REDOX value.

Alarm values
The standard TOPAX® MV 450 measuring amplifiers have two relays, one for min. alarm and one for
max. alarm. To set the limit values, the tipping switch is pushed to the left (min.) or to the right
(max.). The digital display shows the corresponding limit value and can be adjusted at the relevant
potentiometer by means of a screw driver. If the max. value is exceeded or the measured value
remains under the min. limit, one of the relays is activated. The red LED indicates the switching
state of the relay. It it pulls up the LED lights up.